Repeatability at 4.2 K is better than 1mK after 1 year. Accuracy measurement depends on the calibration accuracy and the measurement system accuracy at 10µA measurement current.
For the temperature range 4.2 - 300 K, we normally use 16 - 18 calibration points. The data is supplied in the form of a calibration table and polynomial fit coefficients.
We have recommended mounting methods and can offer advice and support to aid your application of CCS sensors in your systems.
Due to their method of construction, CCS sensors are mechanically robust and bending of the leads is fine providing care is taken. You may solder CCS sensor leads using normal soldering techniques (approx. 250-280oC). As with all delicate instrumentation, we do not recommend heating for longer than 1 min during soldering.
The leads are made from a non-oxidising alloy and as such they are suitable for use at high temperatures. In standard form the sensors are provided with two leads. However, it is possible to modify the exiting leads to give a 4-wire configuration. We can provide an optional design and development service covering applications advice and prototype mounting technology if required.
The sensors are free to the environment, hermetically sealed and are strictly controlled during manufacture.
Calibrated and selected uncalibrated sensors are thermally cycled (50 cycles over 3 months).
Yes. They can be used to measure temperatures in vacuum and in solids, gas and liquids (at pressures up to 25 bar).